BB?00 infant incubator Brief introduction: Simultaneously air temperature and set temperature display. Alarm functions for safety includes Power failure fan failure Over temperature temperature deviation Temperature probe failure >37°C temperature setting. A keypad locks helps to prevent inadvertent changes to settings. Trough for pulling and pushing add the water and wash is convenient. Heater is electrically cut off if air /skin temperature inside incubator exceed 40°C. Humidity is adjustable. The infant bed inclination angle is adjustable.
Technical specification: Electrical requirement : 220V±10%/50±2% Power supply : 420VA Air Temperature control range: 25~38 Temperature fluctuation: 0.5 Uniformity of mattress temperature: 0.8 Warm-up time: 45min Internal noise level: 55dB
Operation condition: Environment temperature: 25~38 Environment relative humidity: 30%~75%RH Environment air velocity of flow: <0.3m/s
Package: Gross weight : Control cabinet: 120kg Hood cabinet: 30kg Net weight: Control cabinet: 90kg Hood cabinet: 10kg Measurement:1200*700*960mm